Philips Avalon FM50 M2705a Philips

$1.00 $0.00


Description

Philips Avalon FM50 Features

The Philips Avalon FM50 fetal/maternal monitors offer a solution for both external and internal fetal monitoring applications, and optional non-invasive maternal vital signs. The Avalon FM50 can monitor fetal heart rates (FHRs) externally using ultrasound, uterine activity using a toco transducer. The Maternal heart rate (MHR) is measured using ECG electrodes. This Philips fetal monitor can also measure Non-Invasive Blood Pressure (NIBP) and maternal oxygen saturation (SpO2). Additionally, the FM50 fetal monitor can monitor one FHR internally via direct fetal electrocardiogram (DECG), and uterine activity internally using an intra-uterine pressure (IUP) catheter. This fetal monitor is capable of intrapartum monitoring.

The measurements are displayed on a 6.5-inch color display as numerics. You can operate the FM50 using the touchscreen interface. Measurements are recorded and can be printed out via the onboard recorder.

This fetal monitor is compatible with the Avalon CTS Cordless Fetal Transducer System. Only one CTS Cordless Fetal Transducer system can be connected at a time. The user has the option to either use wired or cordless fetal transducers. Using both wired and wireless together is not supported in this unit. Cordless transducers have priority over wired transducers. When an Avalon CTS base station is connected via the appropriate interface cable, and there are wired transducers connected, the wired transducers are disabled.

  • Automated coincidence detection with Smart Pulse.
  • The Optional external touch display allows very flexible and ergonomic installations.
  • The Avalon FM series monitors can monitor triplets on a single monitor.
  • Crisp and clear intuitive color touchscreen with large numerics and graphics.
  • Remote display option, including touchscreen capability, for added convenience and safety.
  • Backup memory for seamless monitoring.
  • LAN interface for compatibility with hospital IT networks and access to additional system capabilities.
  • Smart transducers with auto-recognition simplify operation.
  • Transducer plug and play with automatic screen layout to focus on the patient, not the system.
  • Data buffer with multi-patient handling for random printout.
  • Cross-channel verification of maternal and fetal heart rates.

Philips Avalon FM50 Specifications

Dimensions

  • Height: 6.8″ (17.2 cm) ±5%
  • Width: 16.5″ (42 cm) ±5%
  • Depth: 14.6″ (37 cm) ±5%
  • Weight: 19.8 lbs (9 kg)

Power

  • Supply Voltages: 100 VAC to 240 VAC ±10%
  • Supply Frequency Range: 50 Hz/ 50 Hz
  • Power Consumption (Current): 1.3-0.7 A

Degree of Protection Against Electrical Shock

  • Type CF

Electrical Class

  • Class I Equipment

Mode of Operation

  • Continuous Operation

Startup Time

  • Continuous Operation

Temperature Range

Operating

  • 32°F to 113°F (0°C to 45°C)

Storage

  • -4°F to 140°F (-20°C to 60°C)

Humidity Range

Operating

  • < 95% relative humidity @ 40°C/104°F

Storage

  • < 90% relative humidity @ 60°C/140°F

Altitude Range

Operating

  • -500 to 3000 m/-1640 to 9840 ft

Storage

  • -500 to 13100 m/-1640 to 43000 ft.

Philips Avalon FM50 Performance Specifications

Ultrasound

  • Measurement Method: Ultrasound Pulsed Doppler
  • Measurement Range (US): 50 to 240 bpm
  • Resolution
    • Display: 1 bpm
    • Printer: 1/4 bpm
  • Jitter @ 200 bpm: ≤ 2 bpm
  • Display Update Rage: 1/second
  • US Intensity
    • Average output power: P = (4.3 ± 0.4) mW
    • Peak-negative acoustic pressure: p_ = (33.9 ± 3.6) kPa
    • Output beam intensity (Iob) (= spatial average – temporal average intensity): Isata = (2.38 ± 0.75) mW/cm2
    • Spatial-peak temporal average intensity: Ispta = (10.3 ± 2.2) mW/cm2
    • Effective radiating area @ -6 dB: 1.81 cm2
  • Signal Quality Indication
    • Poor: Empty
    • Acceptable: Two-thirds full
    • Good: Full
  • Beat to Beat Change (max.) for Ultrasound: 28 bpm
  • US Frequency: 1 MHz ± 100 Hz
  • US Signal Range: 3.5 μVpp to 350 μVpp @ 200 Hz
  • US Burst
    • Repetition Rate: 3.0 kHz
    • Duration: ≤ 100 μs
  • US LF Frequency Passband @ -3dB: 100 to 500 Hz ± 20%
  • FMP Signal Range @ 33 Hz: 200 μVpp to 40 mVpp
  • FMP Frequency Passband @ -3dB: 10 to 100 Hz

Toco

  • Measurement Method: Strain Gauge Sensor Element
  • Sensitivity: 1 unit = 2.5 g
  • Resolution
    • Display: 1 unit
    • Printer: 1/4 unit
  • Measurement Range: 400 units
  • Signal Range: 0 to 127 Units
  • Maximum Offset Range: -300 units
  • Baseline Setting: 20 units
  • Update Rate
    • Display: 1 / second
    • Printer: ~4 / seconds
  • Auto Offset Correction: 3 seconds after connecting the transducer, the TOCO value is set to 20 units.
  • Auto Zero Adjust: TOCO value is set to zero following a negative measurement value for 5 seconds.

IUP

  • Measurement Method: Passive Resistive Strain Gauge Elements
  • Measurement Range: -100 to +300 mmHg
  • Signal Range: -99 to 127 mmHg
  • Resolution
    • Display: 1 mmHg
    • Printer: 1/4 mmHg
  • Sensitivity: 5 μV/V/mmHg
  • Offset Compensation: +100 to -200 mmHg
  • Baseline Setting: 0 mmHg
  • Accuracy (not including sensor accuracy): ± 0.5% per 100 mmHg
  • Update Rate
    • Display: 1 / second
    • Printer: ~4 / seconds
  • Auto Offset Correction: 3 seconds after connecting the transducer, the IUP value is set to 0 mmHg

 